DS-NH3N200 Ammonia Nitrogen Water Quality Online Automatic Monitor
Contact Info
- Add:西安市高新区锦业二路逸翠尚府北区8号楼1单元1701, Zip: 710000
- Contact: 魏经理
- Tel:15686405979
- Email:1528126689@qq.com
Other Products
Measurement Principle
lUses a sapphire valve core ten-way valve with high wear resistance coefficient, resistance to strong acids and alkalis, organic solvents, long service life, and high stability;
lEmploys a specially designed reference method, where minor displacement of the digestion cell and slight contamination do not affect the measurement value;
lFeatures waste liquid separation function, enabling separate discharge of high-concentration reaction waste and low-concentration cleaning waste;
lHas dynamic dilution function, automatically adjusting the dilution ratio based on the on-site set range;
lIncludes built-in quality control functions such as spiked recovery, blank testing, standard sample verification, zero point verification, span verification, parallel sample testing, and supports automatic, manual, and remote initiation of quality control functions;
lCapable of automatic fault diagnosis and uploading/recording of abnormal information, such as component failure, over-range alarm, exceedance alarm, reagent shortage alarm, etc.;
lEquipped with overheating protection: automatically cuts off power to the heater when the temperature is too high;
lAfter abnormal reset or power restoration, the instrument automatically discharges residual reactants and returns to the initial working state. Features power-on alarm function, generating an alarm message upon restart after power loss;
lHas analysis instrument process log, measurement data recording, and alarm information storage functions. With data recorded hourly, it can store over 3 years of data and supports quick export via USB drive;
lFeatures three-level management permissions;
Typical Applications
lOnline monitoring of ammonia nitrogen concentration in drinking water and surface water;
lOnline monitoring of ammonia nitrogen concentration at municipal sewage inlets/outlets, pollution source discharge points, and industrial wastewater outlets.
Technical Parameters
Range | 0-500mg/L (customizable) |
Indication Error | Within 0-10mg/L range: 20%* ≤±8%; 50%* ≤±5%; 80%* ≤±3% When >10mg/L: ≤±3% |
Repeatability | ≤2% |
Resolution | 0.0001 mg/L (decimal places configurable from 2 to 4 digits) |
Display | 7-inch touch screen |
Measurement Modes | Continuous mode, periodic mode (adjustable from 1h to 30d), fixed-point mode, controlled mode, manual mode optional |
Calibration Modes | Manual calibration and automatic calibration modes |
Maintenance Cycle | Generally once per month |
Power Supply | 220VAC±10%, 50HZ±10% |
Output | 4-20mA output |
Communication Methods | MODBUS RS-232, RS-485 |
Environmental Conditions | Indoor with adjustable temperature, recommended temperature +5~40℃; humidity ≤85% (non-condensing) |
Dimensions | 302*271*462mm (L*W*H) |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 西安东升 |
| Spec: | DS-NH3N200 |
| Stock: | |
| Origin: | China / Shaanxi / Xianshi |